DO METER DO9100
Model: DO9100
DO9100 is a precision meter for measuring dissolved oxygen and temperature. The built-in microcomputer processor can calculate and compensate all measurement data related to dissolved oxygen and temperature. The shell of this instrument is waterproof, and the waterproof grade: IP67
This dissolved oxygen meter has two measurement units: % and mg/L.
The dissolved oxygen probe is a film with a polarographic sensor and a thermistor for temperature measurement and temperature compensation. The permeable film separates the measured solution from the sensing electrode, but allows oxygen to pass through the film to react , The solubility can be measured.
Calibration:
One-point calibration or two-point calibration, any point calibration
One-point calibration: (satisfying normal use) 0.0% (oxygen-free water) or 100.0% (air).
Two-point calibration: 0.0% (oxygen-free water) and 100.0% (air) calibration at the same time
The instrument calibration is simple, make sure the probe is installed correctly before calibration:
1. 100.0%-After the value is stable in the air when the device is turned on, perform calibration. Press and hold the CAL. button for 5 seconds to release the button. The screen flashes 100.0%. After the flashing is completed, the calibration is completed.
2. 0.0%-After the value is stable in the anaerobic water under the power-on state, calibrate, press and hold the CAL. button for 5 seconds to release the button, the screen flashes 0.0%, the flashing is completed, the calibration is completed.
Specifications :
Measuring rangeOxygen concentration (mg/L)0.0-40. 0 mg/L
Dissolved oxygen saturation (%)0.0-300. 0%
Temperature0-40°C; 32-104°F
ResolutionOxygen concentration (mg/L)0.1 mg/L
Dissolved oxygen saturation (%)0.1%
Temperature0, 1°C; 0.1°F
AccuracyOxygen concentration (mg/L)±1.5% °F. S.
Dissolved oxygen saturation (%)±1.5%F. S.
Temperature±0.5℃
Calibration modesupport single-point or dual-point automatic calibration, namely zero point or automatic 100% calibration in the air
Response time≤45s (fast response at 20℃)
Temperature CompensationATC (0-50℃)
Battery powerlow battery warning
Height compensationnone
Shut downautomatically shut down for about 15 minutes
Power supply mode3*1. 5V button battery (not included)
Use environment0-40℃ (32-104°F);RH max 90%
Dimensions and weight183*37*37mm (7.2*1.5*1.5in)/80g